Kumar, K.G. and John, K.S. and lndira, C.J.
  
(2006)
A chloride ion-selective potentiometric sensor based on
a polymeric schiff base complex.
    Indian journal of chemical technology, 13 (1).
     pp. 13-16.
     ISSN 0971-457X
  
  
  
  
  
    
  
    
      
      
    
  
  
  
    Abstract
    This paper describes the fabrication of an ion-selective electrode in which a polymeric schiff basc complex of cobalt(l1)
is used as the ionophore. The main advantage of the electrode is that it is mechanically s~ableu pro 3 nlonths. The clecrrodc
shows a linear response in the range of 2.5 x 10'~-0.5x 10.' mol dm-3.T he response time of the elecrrodc is 30 s. The pH
I-ange at which the electrode works is 3.8 to 6.8. The electrode was found to be selective towards chlol-ide ion in the presence
of ions like Na'. ~ a " .~ n " .~ e ~c'o.2 +,~ i " ,c u2+,z n2'. CH,COO-, NOi, SO?, Brand NO?..
  
  
  
  
  
    Actions (login required)
    
    
      
          | 
        View Item |