Kumar, K.G. and John, K.S. and lndira, C.J. (2006) A chloride ion-selective potentiometric sensor based on a polymeric schiff base complex. Indian journal of chemical technology, 13 (1). pp. 13-16. ISSN 0971-457X

[img]
Preview
PDF - Published Version
Download (279Kb) | Preview

Abstract

This paper describes the fabrication of an ion-selective electrode in which a polymeric schiff basc complex of cobalt(l1) is used as the ionophore. The main advantage of the electrode is that it is mechanically s~ableu pro 3 nlonths. The clecrrodc shows a linear response in the range of 2.5 x 10'~-0.5x 10.' mol dm-3.T he response time of the elecrrodc is 30 s. The pH I-ange at which the electrode works is 3.8 to 6.8. The electrode was found to be selective towards chlol-ide ion in the presence of ions like Na'. ~ a " .~ n " .~ e ~c'o.2 +,~ i " ,c u2+,z n2'. CH,COO-, NOi, SO?, Brand NO?..

Item Type: Article
Uncontrolled Keywords: Ion selective electrodes; Chloride ion; Amino methylated polystyrene