Sathiyanarayanan, S. and Marikkannu, C. and Balasrinivasan, P. and Muthupandi, V. (2002) Corrosion behaviour of Ti6Al4V and duplex stainless steel (UNS 31803) in synthetic bio-fluids. Anti-Corrosion Methods and Materials , 49 (01). pp. 33-37. ISSN 0003-5599
PDF
Restricted to Registered users only Download (4050Kb) | Request a copy |
Abstract
Compares the corrosion behaviour of Ti6Al4V titanium alloy, a conventional duplex stainless steel (UNS 31803) ans AISI 304 austenitic stainless steel in synthetic biofluids using electrochemical techniques and comments on the suitability of DSS for use in biomedical applications. Finds that the general corrosion resistance of duplex stainless steels is slightly inferior to that of austenitic stainless steel and titanium alloy; duplex stainless steel does not show any sign of pitting when exposed to synthetic biofluids and exhibits excellent resistance to localised corrosion on par with that of titanium alloy. Concludes that duplex stainless steels are one of the best alternates to titanium alloys
Item Type: | Article |
---|---|
Uncontrolled Keywords: | Corrosion; Alloys; Steel; Medical products |
Subjects: | Corrosion Science and Engineering |
Divisions: | UNSPECIFIED |
Depositing User: | ttbdu cecri |
Date Deposited: | 08 May 2012 09:22 |
Last Modified: | 08 May 2012 09:22 |
URI: | http://cecri.csircentral.net/id/eprint/2443 |
Actions (login required)
View Item |